Post Offer Free
HORRON XLH Medical Electronics Group  

Gold Index: 4442

You are here: home  > Hospital / clinical Serum Electrolyte Analyzer 60-300ul 220V20V

Hospital / clinical Serum Electrolyte Analyzer 60-300ul 220V20V 

Place of Origin: Zhejiang, China (Mainland) 
inquire
Add to My Favorites
HiSupplier Escrow
Share |

Product Detail

Model No.: IMS-972

 
 
Main Features:
 
 
Since 1997
Simple YES/NO prompted ope

 

 

Main Features:

 

 

Since 1997

Simple YES/NO prompted operation

High accurate and long life electrode and TCO2 sensor

Sleep mode to reduce reagent consumption

Rapid test speed of 80 tests per hour

Minimum sample volume is 60ul

Individual reagent to avoid waste

Option: Auto Sampler

 

 

Parameter configuration:

 

A  K, Na, Cl

B  K, Na, Cl, TCO2

C  K, Na, Cl, iCa, nCa, TCa, pH

D  K, Na, Cl, iCa, nCa, TCa, pH, TCO2, AG  

F  K, Na, Cl, Li  

H  K, Na, Cl, iCa, nCa, TCa, pH, Li

I   K, Na, Cl, iCa, nCa, TCa, pH, Li, TCO2, AG

J   K, Na, Cl, iCa, nCa, TCa, pH, Mg, TCO2, AG

K  K, Na, Cl, iCa, nCa, TCa, pH, Li, Mg, TCO2, AG

 

 

Specifications:

 

Sample: serum, plasma, whole blood, cerebrospinal fluid and dilute urine

Measuring speed: ≤25s

Analysis Method: ion selective electrode (ISE)

Sample volume: 60-300ul

Sampler position: 35 positions (including 1 QC)

Sample tube: Teco cup,

Storage: up to 20,000 test results

Printer: Internal thermal printer

Interface: RS232 port

Working environment:

Temperature: 5-40

Relative Humidity: ≤ 80 %

Atmospheric pressure: (86~106) kPa

Power supply: 220V±20V, 50-60Hz

Power: ≤120W

Dimension: 380mm*270mm*400mm

Net weight: 8Kg

 

 

 

Items

Measuring Range

Resolution

Measuring Precision

(CV%)

K+

0.5-20.0mmol/L

0.01 mmol/L

≤1.0%

Na+

15-200mmol/L

0.1 mmol/L

≤1.0%

Cl-

15-200mmol/L

0.1 mmol/L

≤1.0%

Ca2+

0.1-6.0mmol/L

0.01 mmol/L

≤1.0%

Li+

0.1-5.0mmol/L

0.01 mmol/L

≤2.0%

pH

4-9pH

0.01 pH

≤0.5%

TCO2

2.0-70.0mmol/L

0.1 mmol/L

≤3.0%

 

 

 

Competitive Advantage:

 

 

Simple YES/NO prompted operation

Individual reagent to avoid waste

From 3 up to 11 parameters, with Li, Mg and TCO2 options

16 years proven quality

 

bigPhoto
 Mr. Wang

Tel: 86-20-26855736
Contact to this supplier

Related Search

Find more related products in following catalogs on Hisupplier.com

Related Products

Company Info

HORRON XLH Medical Electronics Group [China (Mainland)]


Business Type:Manufacturer
City: Guangzhou
Province/State: Guangdong
Country/Region: China (Mainland)

You May Like:

Product (94)